Name |
Dissolved Hydrogen Monitor |
Model |
BIH-50D |
Sensor |
LS-1540Y |
Flow Cell |
DH-30FC |
Mearsuring Method |
Membrane Poralographic Cell |
Range |
0.00 - 20.00µg/L (Minimum Indication 0.01μg/L)
0.00 - 10.00 mg/L (Minimum Indication0.01mg/L)
0.0 - 2000 % (Minimum Indication0.1% without analog output) |
Repeatability |
Within ±5%FS (Under fixed conditions) |
Response Time (T90) |
Less than 30 sec. (100%H2 gas → 0%H2gas) |
Indicator |
4-digit LCD |
Analog Output |
4 - 20 mADC, Maximum Load 650Ω |
Power Voltage |
90 - 264VAC 50/60Hz |
Power Comsumption |
Approx. 5W |
Contact Output |
High and Low alarm output: Normal Open
Memtenance-in-progress: Normal Open
Error output, Normal Open |
Contact Capacity |
AC250V 3A DC30V 3A (Resistive load) |
Solution Temperature |
5 - 40°C |
Temperature Compensation Range |
5 - 40°C |
Operating Environment |
5 - 40°C, 20 - 80%Rh (no condensation) |
Fluid Pressure |
Maximum 0.2MPa (2 kgf/cm2) gauge pressure |
Flow Rate |
0.2 - 2.0 L/min. |
Calibration |
Zero calibration – Electrical Zero or pure N2 gas
Span calibration – Atmospere of 100%H2gas |
Instalation Style |
Indoor panel mounting |
Dimensions |
96(W) x 96(H) x 121(D) mm |
Panel Cutting Size |
92(W) x 92(H) mm |
Weight Approx. |
0.5kg |
Digital Communication |
RS-232C (Optional specification) |
Standard Accessaries |
Membrane (M-1540T) : 1 box (10pcs)
Electrolyte (EL-1540T) : 1btl (50mL) |